CVE-2025-54304 - CERT CVE
ID CVE-2025-54304
Sažetak An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
Reference
CVSS
Base: 9.8
Impact: 5.9
Exploitability:3.9
Pristup
VektorSloženostAutentikacija
NETWORK LOW NONE
Impact
PovjerljivostCjelovitostDostupnost
HIGH HIGH HIGH
CVSS vektor CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Zadnje važnije ažuriranje 16-12-2025 - 18:54
Objavljeno 04-12-2025 - 15:15