CVE-2025-4393 - CERT CVE
ID CVE-2025-4393
Sažetak Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Reference
CVSS
Base: 6.5
Impact: 5.5
Exploitability:1.0
Pristup
VektorSloženostAutentikacija
LOCAL HIGH LOW
Impact
PovjerljivostCjelovitostDostupnost
LOW HIGH HIGH
CVSS vektor CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H
Zadnje važnije ažuriranje 25-07-2025 - 15:29
Objavljeno 24-07-2025 - 07:15