CVE-2025-24836 - CERT CVE
ID CVE-2025-24836
Sažetak With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.
Reference
CVSS
Base: 7.1
Impact: 5.5
Exploitability:1.6
Pristup
VektorSloženostAutentikacija
ADJACENT_NETWORK HIGH NONE
Impact
PovjerljivostCjelovitostDostupnost
LOW HIGH HIGH
CVSS vektor CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
Zadnje važnije ažuriranje 13-02-2025 - 22:15
Objavljeno 13-02-2025 - 22:15