| ID |
CVE-2024-9832
|
| Sažetak |
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure. |
| Reference |
|
| CVSS |
| Base: | 9.3 |
| Impact: | 6.0 |
| Exploitability: | 2.5 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| LOCAL |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| HIGH |
HIGH |
HIGH |
|
| CVSS vektor |
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
| Zadnje važnije ažuriranje |
15-11-2024 - 13:58 |
| Objavljeno |
14-11-2024 - 21:15 |