ID |
CVE-2024-48974
|
Sažetak |
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure. |
Reference |
|
CVSS |
Base: | 9.3 |
Impact: | 6.0 |
Exploitability: | 2.5 |
|
Pristup |
Vektor | Složenost | Autentikacija |
LOCAL |
LOW |
NONE |
|
Impact |
Povjerljivost | Cjelovitost | Dostupnost |
HIGH |
HIGH |
HIGH |
|
CVSS vektor |
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
Zadnje važnije ažuriranje |
15-11-2024 - 13:58 |
Objavljeno |
14-11-2024 - 22:15 |