ID |
CVE-2024-48967
|
Sažetak |
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance. |
Reference |
|
CVSS |
Base: | 10.0 |
Impact: | 6.0 |
Exploitability: | 3.9 |
|
Pristup |
Vektor | Složenost | Autentikacija |
NETWORK |
LOW |
NONE |
|
Impact |
Povjerljivost | Cjelovitost | Dostupnost |
HIGH |
HIGH |
HIGH |
|
CVSS vektor |
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
Zadnje važnije ažuriranje |
15-11-2024 - 13:58 |
Objavljeno |
14-11-2024 - 22:15 |