| ID |
CVE-2023-47262
|
| Sažetak |
The startup process and device configurations of the Abbott ID NOW device, before v7.1, can be interrupted and/or modified via physical access to an internal serial port. Direct physical access is required to exploit. |
| Reference |
|
| CVSS |
| Base: | 5.2 |
| Impact: | 4.2 |
| Exploitability: | 0.9 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| PHYSICAL |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| NONE |
HIGH |
LOW |
|
| CVSS vektor |
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:L |
| Zadnje važnije ažuriranje |
13-12-2023 - 07:15 |
| Objavljeno |
14-11-2023 - 16:15 |