| ID |
CVE-2021-33882
|
| Sažetak |
A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands. |
| Reference |
|
| CVSS |
| Base: | 5.0 |
| Impact: | 2.9 |
| Exploitability: | 10.0 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| NONE |
PARTIAL |
NONE |
|
| CVSS vektor |
AV:N/AC:L/Au:N/C:N/I:P/A:N |
| Zadnje važnije ažuriranje |
01-09-2021 - 15:52 |
| Objavljeno |
25-08-2021 - 12:15 |