| ID |
CVE-2020-15486
|
| Sažetak |
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity. |
| Reference |
|
| CVSS |
| Base: | 3.3 |
| Impact: | 2.9 |
| Exploitability: | 6.5 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| ADJACENT_NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| PARTIAL |
NONE |
NONE |
|
| CVSS vektor |
AV:A/AC:L/Au:N/C:P/I:N/A:N |
| Zadnje važnije ažuriranje |
21-07-2021 - 11:39 |
| Objavljeno |
26-08-2020 - 16:15 |