ID |
CVE-2020-12037
|
Sažetak |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. |
Reference |
|
CVSS |
Base: | 5.0 |
Impact: | 2.9 |
Exploitability: | 10.0 |
|
Pristup |
Vektor | Složenost | Autentikacija |
NETWORK |
LOW |
NONE |
|
Impact |
Povjerljivost | Cjelovitost | Dostupnost |
PARTIAL |
NONE |
NONE |
|
CVSS vektor |
AV:N/AC:L/Au:N/C:P/I:N/A:N |
Zadnje važnije ažuriranje |
14-07-2020 - 15:41 |
Objavljeno |
29-06-2020 - 14:15 |