CVE-2019-25723 - CERT CVE
ID CVE-2019-25723
Sažetak Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Reference
CVSS
Base: 4.0
Impact: 1.4
Exploitability:2.2
Pristup
VektorSloženostAutentikacija
NETWORK HIGH NONE
Impact
PovjerljivostCjelovitostDostupnost
NONE NONE LOW
CVSS vektor CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Zadnje važnije ažuriranje 02-06-2026 - 20:16
Objavljeno 02-06-2026 - 20:16