| ID |
CVE-2019-25722
|
| Sažetak |
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring. |
| Reference |
|
| CVSS |
| Base: | 7.6 |
| Impact: | 4.7 |
| Exploitability: | 2.8 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| ADJACENT_NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| LOW |
LOW |
HIGH |
|
| CVSS vektor |
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| Zadnje važnije ažuriranje |
02-06-2026 - 20:16 |
| Objavljeno |
02-06-2026 - 20:16 |