| ID |
CVE-2019-25721
|
| Sažetak |
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality. |
| Reference |
|
| CVSS |
| Base: | 6.5 |
| Impact: | 3.6 |
| Exploitability: | 2.8 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| ADJACENT_NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| NONE |
NONE |
HIGH |
|
| CVSS vektor |
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
| Zadnje važnije ažuriranje |
02-06-2026 - 20:16 |
| Objavljeno |
02-06-2026 - 20:16 |