| ID |
CVE-2019-25719
|
| Sažetak |
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality. |
| Reference |
|
| CVSS |
| Base: | 8.6 |
| Impact: | 4.7 |
| Exploitability: | 3.9 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| LOW |
LOW |
HIGH |
|
| CVSS vektor |
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| Zadnje važnije ažuriranje |
02-06-2026 - 14:40 |
| Objavljeno |
02-06-2026 - 14:16 |