| ID |
CVE-2019-18248
|
| Sažetak |
BIOTRONIK CardioMessenger II, The affected products transmit credentials in clear-text prior to switching to an encrypted communication channel. An attacker can disclose the product’s client credentials for connecting to the BIOTRONIK Remote Communication infrastructure. |
| Reference |
|
| CVSS |
| Base: | 3.3 |
| Impact: | 2.9 |
| Exploitability: | 6.5 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| ADJACENT_NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| PARTIAL |
NONE |
NONE |
|
| CVSS vektor |
AV:A/AC:L/Au:N/C:P/I:N/A:N |
| Zadnje važnije ažuriranje |
06-04-2021 - 17:16 |
| Objavljeno |
29-06-2020 - 14:15 |