CVE-2018-10601 - CERT CVE
ID CVE-2018-10601
Sažetak IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Reference
CVSS
Base: 5.4
Impact: 6.4
Exploitability:5.5
Pristup
VektorSloženostAutentikacija
ADJACENT_NETWORK MEDIUM NONE
Impact
PovjerljivostCjelovitostDostupnost
PARTIAL PARTIAL PARTIAL
CVSS vektor AV:A/AC:M/Au:N/C:P/I:P/A:P
Zadnje važnije ažuriranje 10-05-2021 - 15:08
Objavljeno 05-06-2018 - 20:29