CVE-2017-12724 - CERT CVE
ID CVE-2017-12724
Sažetak A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
Reference
CVSS
Base: 6.8
Impact: 6.4
Exploitability:8.6
Pristup
VektorSloženostAutentikacija
NETWORK MEDIUM NONE
Impact
PovjerljivostCjelovitostDostupnost
PARTIAL PARTIAL PARTIAL
CVSS vektor AV:N/AC:M/Au:N/C:P/I:P/A:P
Zadnje važnije ažuriranje 02-03-2018 - 14:40
Objavljeno 15-02-2018 - 10:29