ID |
CVE-2017-12714
|
Sažetak |
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities. |
Reference |
|
CVSS |
Base: | 6.1 |
Impact: | 6.9 |
Exploitability: | 6.5 |
|
Pristup |
Vektor | Složenost | Autentikacija |
ADJACENT_NETWORK |
LOW |
NONE |
|
Impact |
Povjerljivost | Cjelovitost | Dostupnost |
NONE |
NONE |
COMPLETE |
|
CVSS vektor |
AV:A/AC:L/Au:N/C:N/I:N/A:C |
Zadnje važnije ažuriranje |
09-10-2019 - 23:23 |
Objavljeno |
25-04-2018 - 13:29 |