| ID |
CVE-2015-3954
|
| Sažetak |
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue. |
| Reference |
|
| CVSS |
| Base: | 10.0 |
| Impact: | 10.0 |
| Exploitability: | 10.0 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| COMPLETE |
COMPLETE |
COMPLETE |
|
| CVSS vektor |
AV:N/AC:L/Au:N/C:C/I:C/A:C |
| Zadnje važnije ažuriranje |
09-10-2019 - 23:14 |
| Objavljeno |
25-03-2019 - 17:29 |