CVE-2015-3952
|
| Sažetak |
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue. |
| Reference |
|
| CVSS |
| Base: | 5.0 |
| Impact: | 2.9 |
| Exploitability: | 10.0 |
|
| Pristup |
| Vektor | Složenost | Autentikacija |
| NETWORK |
LOW |
NONE |
|
| Impact |
| Povjerljivost | Cjelovitost | Dostupnost |
| PARTIAL |
NONE |
NONE |
|
| CVSS vektor |
AV:N/AC:L/Au:N/C:P/I:N/A:N |
| Zadnje važnije ažuriranje |
09-10-2019 - 23:14 |