ID |
CVE-2014-5434
|
Sažetak |
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. |
Reference |
|
CVSS |
Base: | 5.0 |
Impact: | 2.9 |
Exploitability: | 10.0 |
|
Pristup |
Vektor | Složenost | Autentikacija |
NETWORK |
LOW |
NONE |
|
Impact |
Povjerljivost | Cjelovitost | Dostupnost |
PARTIAL |
NONE |
NONE |
|
CVSS vektor |
AV:N/AC:L/Au:N/C:P/I:N/A:N |
Zadnje važnije ažuriranje |
09-10-2019 - 23:11 |
Objavljeno |
26-03-2019 - 15:29 |